Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractEvaluation of an air sampling technique for assessing low-level volatile organic compounds in indoor environments    Next AbstractExposure to volatile organic compounds in healthcare settings »

J Air Waste Manag Assoc


Title:Effect of calibration environment on the performance of direct-reading organic vapor monitors
Author(s):LeBouf RF; Slaven JE; Coffey CC;
Address:"Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, 1095 Willowdale Road, Morgantown, WV 26505, USA. rlebouf@cdc.gov"
Journal Title:J Air Waste Manag Assoc
Year:2013
Volume:63
Issue:5
Page Number:528 - 533
DOI: 10.1080/10962247.2013.772926
ISSN/ISBN:1096-2247 (Print) 1096-2247 (Linking)
Abstract:"The performance of two direct-reading organic vapor monitors (monitors) when calibrated at different environmental conditions was compared with charcoal tube results. Three MIRAN SapphIRe portable ambient air analyzers (SAP) and three Century portable toxic vapor analyzers (TVAs) were evaluated. Prior to sampling, the monitors were calibrated per the manufacturer's instructions using methane for the TVA flame ionization detector (FID) and isobutylene for the photoionization detector (PID), whereas the SapphIRe instruments were zeroed and the instrument's manufacturer-supplied library was used. For the first series of tests ('Part 1--Same condition'), the monitors were calibrated under the same environmental conditions as those present during sampling. They were then challenged with four cyclohexane concentrations (30, 150, 300, and 475 ppm) under two extreme environmental conditions: 5 degrees C and 30% relative humidity (RH) (same/cold) and 38 degrees C and 90% RH (same/hot). For the second series of tests ('Part 2--Different condition'), the monitors were calibrated at approximately normal indoor environmental conditions (21 degrees C and 50% RH) and sampled at extreme environmental conditions (different/cold and different/hot). The monitor readings from the two methods were compared with the actual cyclohexane concentration determined from charcoal tubes using ratios and root mean square errors. A number of monitor failures, both below detection limit values in the presence of a known challenge concentration and erroneously high measurements, occurred in each part: same condition 20.7% (149/720) and different condition 42.4% (305/ 720), with a majority of the failures (> 78%) during the hot and humid conditions. All monitors performed best at the same/cold, followed by the same/hot, in terms of closeness to the reference standard method and low within-monitor variability. The ranked choice of monitors for same/cold is PID > SAP > FID; for different/cold FID > PID > SAP; for same/hot SAP > PID > FID; and for different/hot PID > SAP (FID not included due to 100% failure rate). IMPLICATIONS: Direct-reading organic vapor monitors are used for assessing the concentrations of volatile organic compounds in the air at varying environmental conditions. Typical calibration is performed at laboratory temperature and pressure. The monitors may be used in atmospheres that differ from that during calibration. An understanding of the effect of calibration environment on monitor performance may provide valuable information on the reliability and appropriateness of certain monitor types for industrial hygienists, emergency responders, and exposure assessment practitioners. Results of the study indicate monitor calibration should be performed at the same environmental conditions as sampling"
Keywords:Air Pollutants/*analysis Calibration Environmental Monitoring/*instrumentation Gases/*analysis Humidity Temperature;
Notes:"MedlineLeBouf, Ryan F Slaven, James E Coffey, Christopher C eng Comparative Study Research Support, U.S. Gov't, P.H.S. 2013/06/22 J Air Waste Manag Assoc. 2013 May; 63(5):528-33. doi: 10.1080/10962247.2013.772926"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 04-12-2024