« Previous AbstractCavitand-Decorated Silicon Columnar Nanostructures for the Surface Recognition of Volatile Nitroaromatic Compounds    Next AbstractNon-volatile and volatile composition of West African bulk and Ecuadorian fine-flavor cocoa liquor and chocolate »

J Chromatogr A


Title:Dynamic versus static sampling for the quantitative analysis of volatile organic compounds in air with polydimethylsiloxane-carboxen solid-phase microextraction fibers
Author(s):Tuduri L; Desauziers V; Fanlo JL;
Address:"Laboratoire Genie de l'Environnement Industriel, Ecole des Mines d'Ales, France. ludovic.tuduri@ema.fr"
Journal Title:J Chromatogr A
Year:2002
Volume:963
Issue:1-Feb
Page Number:49 - 56
DOI: 10.1016/s0021-9673(02)00222-4
ISSN/ISBN:0021-9673 (Print) 0021-9673 (Linking)
Abstract:"Polydimethylsiloxane-Carboxen solid-phase microextraction fibers are now well known to be very efficient trapping media for the analysis of volatile organic compound (VOC) traces in air. However, competitive adsorption, due to the nature of the coating, considerably limits analyte quantitation. In this contribution, different experimental conditions are investigated to achieve quantitative analysis. Static and dynamic sampling were compared for the analysis of 11 VOCs in a standard gaseous mixture at different extraction times (1, 5, 15 and 45 min). The same experiments were performed with four isolated compounds. Adsorption results from gas mixture and isolated compounds were compared and a common linear range (i.e., where quantitative analysis is conceivable) was determined. When sampling was in the dynamic mode, compounds with lower affinity for the coating showed a very narrow linear range, meaning that competition for adsorption was quickly discriminative. The same experiments in static mode allowed one to obtain wider linear ranges for all compounds, especially for lower-affinity compounds: for a 1 min sampling time, acetone showed a linear adsorption range from 3 to 60 microg m(-3) in the dynamic mode which extended from 5 to 300 microg m(-3) in the static mode"
Keywords:Air/*analysis Calibration Dimethylpolysiloxanes/*chemistry Organic Chemicals/*analysis Reproducibility of Results Sensitivity and Specificity Silicones/*chemistry Volatilization;
Notes:"MedlineTuduri, Ludovic Desauziers, Valerie Fanlo, Jean Louis eng Netherlands 2002/08/22 J Chromatogr A. 2002 Jul 19; 963(1-2):49-56. doi: 10.1016/s0021-9673(02)00222-4"

 