Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractChemotherapy control by breath profile with application of SPME-GC/MS method    Next Abstract[Consumption of volatile organic compounds by alcaliphilic microorganisms] »

J Chromatogr Sci


Title:Evaluation of septa quality for automatic SPME-GC-MS trace analysis
Author(s):Ulanowska A; Ligor T; Amann A; Buszewski B;
Address:"Chair of Environmental Chemistry and Bioanalytics, Faculty of Chemistry, Nicolaus Copernicus University Gagarin Street 7, 87-100 Torun, Poland"
Journal Title:J Chromatogr Sci
Year:2012
Volume:50
Issue:1
Page Number:10 - 14
DOI: 10.1093/chromsci/bmr006
ISSN/ISBN:1945-239X (Electronic) 0021-9665 (Print) 0021-9665 (Linking)
Abstract:"The vials used for the preparation of breath samples for automated solid-phase microextraction-gas chromatography-mass spectrometry analysis are crimped with septa. These septa often emit specific volatile organic compounds (VOCs) confounding the measurement results of breath samples. In the current paper, 14 different brands of magnetic caps with silicone-polytetrafluoroethylene (PTFE), butyl-PTFE, or butyl rubber septa were tested. The total emission of septa over a 4 h period was also evaluated. The tested septa emitted 39 different compounds, which are mainly hydrocarbons, alcohols, and ketones. Acetone and toluene are the most abundant out-gassing products. The concentration of acetone was in the range from 55 to 694 ppb for butyl-PTFE septum (brand 14) and butyl rubber (brand 10), respectively. The measured toluene amount was 69-1323 ppb for the septum brand 14 and brand 8 (silicone-PTFE), respectively. Generally, the butyl rubber septa released higher amounts of contaminants in comparison to the silicone ones"
Keywords:Acetone/analysis/chemistry Breath Tests/*instrumentation Gas Chromatography-Mass Spectrometry/*instrumentation Humans Polytetrafluoroethylene/chemistry Reproducibility of Results Rubber/chemistry Silicones/chemistry Solid Phase Microextraction/*instrument;
Notes:"MedlineUlanowska, Agnieszka Ligor, Tomasz Amann, Anton Buszewski, Boguslaw eng Research Support, Non-U.S. Gov't 2012/02/01 J Chromatogr Sci. 2012 Jan; 50(1):10-4. doi: 10.1093/chromsci/bmr006"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 27-12-2024