Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous Abstract"Evaluation of microfiltration and heat treatment on the microbiological characteristics, phenolic composition and volatile compound profile of pomegranate (Punica granatum L.) juice"    Next AbstractInsect oviposition induces volatile emission in herbaceous plants that attracts egg parasitoids »

J Perinatol


Title:Emission of volatile organic compounds from medical equipment inside neonatal incubators
Author(s):Colareta Ugarte U; Prazad P; Puppala BL; Schweig L; Donovan R; Cortes DR; Gulati A;
Address:"Division of Neonatology, Department of Pediatrics, Advocate Lutheran General Hospital/Advocate Children's Hospital, Park Ridge, IL, USA. 1] Division of Neonatology, Department of Pediatrics, Advocate Lutheran General Hospital/Advocate Children's Hospital, Park Ridge, IL, USA [2] Advocate Medical Group, Park Ridge, IL, USA. Division of Research, Department of Pediatrics, Advocate Lutheran General Hospital/Advocate Children's Hospital, Park Ridge, IL, USA. STAT Analysis Corporation, Chicago, IL, USA. Chicago College of Pharmacy, Midwestern University, Downers Grove, IL, USA"
Journal Title:J Perinatol
Year:2014
Volume:20140424
Issue:8
Page Number:624 - 628
DOI: 10.1038/jp.2014.65
ISSN/ISBN:1476-5543 (Electronic) 0743-8346 (Linking)
Abstract:"OBJECTIVE: To determine emission of volatile organic compounds (VOCs) from plastic medical equipment within an incubator. STUDY DESIGN: Air samples from incubators before and after adding medical equipment were analyzed using EPA TO-15 methodology. Headspace analysis was used to identify VOC emissions from each medical equipment item. Air changes per hour (ACH) of each incubator were determined and used to calculate the emission rate of identified VOCs. RESULTS: Cyclohexanone was identified in all incubator air samples. At 28 degrees C, the mean concentration before and after adding medical equipment items was 2.1 +/- 0.6 and 57.2 +/- 14.9 mug m(-3),respectively (P<0.01). Concentrations increased to a mean of 83.8 +/- 23.8 mug m(-)(3) (P<0.01) at 37(o)C and 93.0 +/- 45.1 mug m(-)(3) (P=0.39) after adding 50% humidity. Intravenous tubing contributed 89% of cyclohexanone emissions. ACH were determined with access doors closed and open with means of 11.5 +/- 1.7 and 44.1 +/- 6.7 h(-1), respectively. Cyclohexanone emission rate increased from a mean of 102.2 mug h(-1) at 28( degrees C to 148.8 mug h(-1) (P<0.01) at 37 degrees C. CONCLUSION: Cyclohexanone was quantified in all incubator air samples containing plastic medical equipment. The concentration of cyclohexanone within the incubator was inversely related to ACH in the closed mode. The cyclohexanone concentration as well as the emission rate increased with higher temperature"
Keywords:"Air Pollution, Indoor/*analysis Bedding and Linens Beds Environmental Exposure/*analysis Environmental Monitoring *Incubators, Infant Plastics Volatile Organic Compounds/*analysis;"
Notes:"MedlineColareta Ugarte, U Prazad, P Puppala, B L Schweig, L Donovan, R Cortes, D R Gulati, A eng Research Support, Non-U.S. Gov't 2014/04/26 J Perinatol. 2014 Aug; 34(8):624-8. doi: 10.1038/jp.2014.65. Epub 2014 Apr 24"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 23-09-2024