Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractChemoreceptors in lepidoptera: stereochemical differentiation of dual receptors for an achiral pheromone    Next AbstractContact chemoreception in feeding by phytophagous insects »

Anal Chem


Title:Facile hyphenation of gas chromatography and a microcantilever array sensor for enhanced selectivity
Author(s):Chapman PJ; Vogt F; Dutta P; Datskos PG; Devault GL; Sepaniak MJ;
Address:"Department of Chemistry, University of Tennessee, Knoxville, Tennessee 37996-1600, USA"
Journal Title:Anal Chem
Year:2007
Volume:79
Issue:1
Page Number:364 - 370
DOI: 10.1021/ac061389x
ISSN/ISBN:0003-2700 (Print) 0003-2700 (Linking)
Abstract:"The very simple coupling of a standard, packed-column gas chromatograph with a microcantilever array (MCA) is demonstrated for enhanced selectivity and potential analyte identification in the analysis of volatile organic compounds (VOCs). The cantilevers in MCAs are differentially coated on one side with responsive phases (RPs) and produce bending responses of the cantilevers due to analyte-induced surface stresses. Generally, individual components are difficult to elucidate when introduced to MCA systems as mixtures, although pattern recognition techniques are helpful in identifying single components, binary mixtures, or composite responses of distinct mixtures (e.g., fragrances). In the present work, simple test VOC mixtures composed of acetone, ethanol, and trichloroethylene (TCE) in pentane and methanol and acetonitrile in pentane are first separated using a standard gas chromatograph and then introduced into a MCA flow cell. Significant amounts of response diversity to the analytes in the mixtures are demonstrated across the RP-coated cantilevers of the array. Principal component analysis is used to demonstrate that only three components of a four-component VOC mixture could be identified without mixture separation. Calibration studies are performed, demonstrating a good linear response over 2 orders of magnitude for each component in the primary study mixture. Studies of operational parameters including column temperature, column flow rate, and array cell temperature are conducted. Reproducibility studies of VOC peak areas and peak heights are also carried out showing RSDs of less than 4 and 3%, respectively, for intra-assay studies. Of practical significance is the facile manner by which the hyphenation of a mature separation technique and the burgeoning sensing approach is accomplished, and the potential to use pattern recognition techniques with MCAs as a new type of detector for chromatography with analyte-identifying capabilities"
Keywords:"Acetone/analysis Acetonitriles/analysis Biosensing Techniques/*methods Chromatography, Gas/*methods Ethanol/analysis Methanol/analysis Organic Chemicals/*analysis Pentanes/*analysis/*chemistry Principal Component Analysis Reproducibility of Results Sensit;"
Notes:"MedlineChapman, Peter J Vogt, Frank Dutta, Pampa Datskos, Panos G Devault, Gerald L Sepaniak, Michael J eng Research Support, Non-U.S. Gov't 2006/12/30 Anal Chem. 2007 Jan 1; 79(1):364-70. doi: 10.1021/ac061389x"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 27-12-2024