Title: | Optimized Method for Determination 57 Volatile Organic Compounds in Nitrogen Using GC x GC-FID |
Author(s): | Yang J; Du J; Huang W; Ning H; Li N; |
Address: | "Institute for Environmental Reference Materials, Environmental Development Center of the Ministry of Ecology and Environment, No. 1 Yuhui Road, Chaoyang, Beijing, 10029, China" |
ISSN/ISBN: | 1945-239X (Electronic) 0021-9665 (Linking) |
Abstract: | "The sample containing 57 volatile organic compounds (VOCs) in nitrogen at a nominal 1 ppmv was prepared in our lab using weighting method. A methodology for determination of the 57 VOCs using a two-dimensional gas chromatography equipped with Deans switches and two flame ionization detectors (GC x GC-FID) was developed and validated for resolution, asymmetry, sensitivity, precision (intra-day precision and inter-day precision), linearity, limit of detection (LOD), limit of quantification (LOQ) and accuracy. In this study, resolution, asymmetry and sensitivity of the analytical method were improved??eintra-day precisions of all the compounds were <1% and inter-day precisions were between 0.9 and 3.0%. In addition, LOQ and LOD were in the range of 0.024-0.185 ppmv and 0.012-0.092 ppmv, respectively. An excellent linearity was obtained (R2 > 0.9995). At the meantime, the accuracy of the analytical method was evaluated by determining the concentration of a certified reference material" |
Keywords: | "Chromatography, Gas/methods Flame Ionization/methods Limit of Detection Nitrogen *Volatile Organic Compounds/analysis;" |
Notes: | "MedlineYang, Jing Du, Jian Huang, Wei Ning, Hongbing Li, Ning eng 2016-16/Ministry of Ecology and Environment of the People's Republic of China/ 2021/10/24 J Chromatogr Sci. 2022 Oct 3; 60(8):713-724. doi: 10.1093/chromsci/bmab118" |