Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractCharacterization of the Sensory Properties and Quality Components of Huangjin Green Tea Based on Molecular Sensory-Omics    Next AbstractFluorescence competition assay for the assessment of green leaf volatiles and trans-beta-farnesene bound to three odorant-binding proteins in the wheat aphid Sitobion avenae (Fabricius) »

Analyst


Title:Characterization of a high-performance portable GC with a chemiresistor array detector
Author(s):Zhong Q; Steinecker WH; Zellers ET;
Address:"Department of Environmental Health Sciences, University of Michigan, 109 S. Observatory, Ann Arbor, MI 48109-2029, USA"
Journal Title:Analyst
Year:2009
Volume:20081112
Issue:2
Page Number:283 - 293
DOI: 10.1039/b810944c
ISSN/ISBN:1364-5528 (Electronic) 0003-2654 (Linking)
Abstract:"The laboratory characterization of a novel, second-generation portable gas chromatograph (GC) prototype designed for trace-level determinations of complex mixtures of volatile organic compounds (VOC) is described. The instrument incorporates a small, multi-stage adsorbent preconcentrator/injector (PCI), two series-coupled separation columns with fast, independent temperature-programming capabilities and junction-point pressure/flow control, and a detector consisting of an array of microfabricated chemiresistor (CR) sensors coated with thiolate-monolayer-protected gold nanoparticle films. Response patterns from the CR array are used in conjunction with chromatographic retention times to identify eluting mixture components. Scrubbed ambient air is used as the carrier gas. Enhancements in design relative to a previously reported first-generation prototype instrument have led to significant reductions in limits of detection as well as improvements in resolution, reliability, flexibility, and convenience. Key features of the instrument are characterized, with an emphasis on the tradeoffs in sensor array performance associated with operation at different temperatures and flow rates. The separation of a preconcentrated mixture of 31 VOCs in < 7 minutes is demonstrated. Projected detection limits are in the ppt range for most compounds, assuming a 1 L sample volume"
Keywords:"Air/*analysis Air Pollutants/*analysis Chromatography, Gas/*instrumentation/methods Environmental Monitoring/*instrumentation/methods Equipment Design Volatile Organic Compounds/*analysis;"
Notes:"MedlineZhong, Qiongyan Steinecker, William H Zellers, Edward T eng R01-OH03692/OH/NIOSH CDC HHS/ Research Support, U.S. Gov't, Non-P.H.S. Research Support, U.S. Gov't, P.H.S. England 2009/01/29 Analyst. 2009 Feb; 134(2):283-93. doi: 10.1039/b810944c. Epub 2008 Nov 12"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 26-12-2024