Title: | Demonstration of proton-transfer reaction time-of-flight mass spectrometry for real-time analysis of trace volatile organic compounds |
Author(s): | Blake RS; Whyte C; Hughes CO; Ellis AM; Monks PS; |
Address: | "Department of Chemistry, University of Leicester, Leicester LE1 7RH, U.K" |
ISSN/ISBN: | 0003-2700 (Print) 0003-2700 (Linking) |
Abstract: | "A proton-transfer reaction mass spectrometer based on time-of-flight mass spectrometry is described. This instrument couples a radioactive ion source and drift tube with a reflectron time-of-flight mass spectrometer. Volatile organic compounds in the gas phase with concentrations at the parts per billion by volume level can be detected in a matter of seconds, and crucially, the multichannel data acquisition in TOF-MS means that this detection sensitivity is available in all mass channels simultaneously. The typical mass resolution (m/Deltam) is in excess of 1000. The performance of the instrument is demonstrated using urban air measurements and a linear response/calibration test" |
Keywords: | Air Pollutants/*analysis Calibration Hydrogen Sulfide/*analysis Linear Models Mass Spectrometry/*instrumentation/*methods Organic Chemicals/*analysis Protons Sensitivity and Specificity Surface Properties Volatilization Water/chemistry; |
Notes: | "MedlineBlake, Robert S Whyte, Christopher Hughes, Ceri O Ellis, Andrew M Monks, Paul S eng Research Support, Non-U.S. Gov't 2004/07/02 Anal Chem. 2004 Jul 1; 76(13):3841-5. doi: 10.1021/ac0498260" |