Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractIdentification of a damage-associated molecular pattern (DAMP) receptor and its cognate peptide ligand in sweet potato (Ipomoea batatas)    Next AbstractAnalysis of chemical compositions contributable to chemical oxygen demand (COD) of oilfield produced water »

Sensors (Basel)


Title:A Microvalve Module with High Chemical Inertness and Embedded Flow Heating for Microscale Gas Chromatography
Author(s):Lu HT; Qin Y; Gianchandani Y;
Address:"Center for Wireless Integrated MicroSensing and Systems (WIMS2), University of Michigan, Ann Arbor, MI 48109, USA. Department of Mechanical Engineering, University of Michigan, Ann Arbor, MI 48109, USA. Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI 48109, USA"
Journal Title:Sensors (Basel)
Year:2021
Volume:20210118
Issue:2
Page Number: -
DOI: 10.3390/s21020632
ISSN/ISBN:1424-8220 (Electronic) 1424-8220 (Linking)
Abstract:"This paper reports a multi-valve module with high chemical inertness and embedded flow heating for microscale gas chromatography (microGC) systems. The multi-valve module incorporates a monolithically microfabricated die stack, polyimide valve membranes, and solenoid actuators. The design incorporates three valves within a single module of volume 30.2 cm(3), which is suitable for the small form factor of microGC systems. The die stack uses fused silica wafers and polyimide valve membranes that enhance chemical inertness. The monolithic die stack requires only three lithographic masks to pattern fluidic microchannels, valve seats, and thin-film metal heaters and thermistors. The performance of fabricated multi-valve modules is compared to a commercial valve in tests using multiple volatile organic compounds, including alkanes, alcohols, ketones, aromatic hydrocarbons, and phosphonates. The valves show almost no distortion of chromatographic peaks. The experimentally measured ratio of flow conductance is 3.46 x 10(3), with 4.15 sccm/kPa in the open state and 0.0012 sccm/kPa in the closed state. The response time is <120 ms"
Keywords:chemical warfare agents organophosphorus compounds phosphonates solenoid volatile organic compounds;
Notes:"PubMed-not-MEDLINELu, Hsueh-Tsung Qin, Yutao Gianchandani, Yogesh eng FA8650-17-C-9105 and FA8650-19-C-9105/Intelligence Advanced Research Projects Activity/ Switzerland 2021/01/23 Sensors (Basel). 2021 Jan 18; 21(2):632. doi: 10.3390/s21020632"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 22-11-2024