Title: | Detector for liquid chromatography based on acoustic emissions from an oscillating flame |
Address: | "Department of Chemistry, University of Calgary, 2500 University Drive, N.W., Calgary, Alberta, Canada T2N 1N4. thurbide@ucalgary.ca" |
ISSN/ISBN: | 0003-2700 (Print) 0003-2700 (Linking) |
Abstract: | "The acoustic flame detector (AFD) is examined as a novel detector for liquid chromatography (LC). It is based upon the acoustic emission frequency of an oscillating hydrogen/oxygen premixed flame and produces a universal response toward organic molecules. A stable frequency near 1000 Hz, which further depends on mobile-phase composition, is achieved for flow rates in the microliter per minute range. The mass flow sensitivity of the AFD demonstrates a linear response over 3 orders of magnitude and a detection limit (S/sigma = 3) of approximately 15 ng of C/s for a series of alcohols. For cyclopentanol, this amounts to an injected mass of approximately 77 ng based on a 0.5-microL injection of a 196 ppm solution in methanol (flow rate 20 microL/min methanol; peak width 30 s). Similar sensitivity is observed using a water mobile phase. Low-frequency (1/f ) noise contributions are dominant with or without mobile phase present. The AFD demonstrates a uniform molar sensitivity toward carbon compounds independent of their optical properties or volatility. Results suggest the device might serve as a simple, inexpensive universal LC detector" |
Notes: | "PubMed-not-MEDLINEThurbide, Kevin B Xia, Zhongpeng eng 2004/09/15 Anal Chem. 2004 Sep 15; 76(18):5459-64. doi: 10.1021/ac049777r" |