Title: | Mercury Sorption and Desorption on Gold: A Comparative Analysis of Surface Acoustic Wave and Quartz Crystal Microbalance-Based Sensors |
Author(s): | Kabir KM; Sabri YM; Esmaielzadeh Kandjani A; Matthews GI; Field M; Jones LA; Nafady A; Ippolito SJ; Bhargava SK; |
Address: | "parallelDepartment of Chemistry, Faculty of Science, Sohag University, Sohag, Egypt. perpendicularDepartment of Chemistry, College of Science, King Saud University, Riyadh, Saudi Arabia" |
DOI: | 10.1021/acs.langmuir.5b01858 |
ISSN/ISBN: | 1520-5827 (Electronic) 0743-7463 (Linking) |
Abstract: | "Microelectromechanical sensors based on surface acoustic wave (SAW) and quartz crystal microbalance (QCM) transducers possess substantial potential as online elemental mercury (Hg(0)) vapor detectors in industrial stack effluents. In this study, a comparison of SAW- and QCM-based sensors is performed for the detection of low concentrations of Hg(0) vapor (ranging from 24 to 365 ppbv). Experimental measurements and finite element method (FEM) simulations allow the comparison of these sensors with regard to their sensitivity, sorption and desorption characteristics, and response time following Hg(0) vapor exposure at various operating temperatures ranging from 35 to 75 degrees C. Both of the sensors were fabricated on quartz substrates (ST and AT cut quartz for SAW and QCM devices, respectively) and employed thin gold (Au) layers as the electrodes. The SAW-based sensor exhibited up to approximately 111 and approximately 39 times higher response magnitudes than did the QCM-based sensor at 35 and 55 degrees C, respectively, when exposed to Hg(0) vapor concentrations ranging from 24 to 365 ppbv. The Hg(0) sorption and desorption calibration curves of both sensors were found to fit well with the Langmuir extension isotherm at different operating temperatures. Furthermore, the Hg(0) sorption and desorption rate demonstrated by the SAW-based sensor was found to decrease as the operating temperature increased, while the opposite trend was observed for the QCM-based sensor. However, the SAW-based sensor reached the maximum Hg(0) sorption rate faster than the QCM-based sensor regardless of operating temperature, whereas both sensors showed similar response times (t90) at various temperatures. Additionally, the sorption rate data was utilized in this study in order to obtain a faster response time from the sensor upon exposure to Hg(0) vapor. Furthermore, comparative analysis of the developed sensors' selectivity showed that the SAW-based sensor had a higher overall selectivity (90%) than did the QCM counterpart (84%) while Hg(0) vapor was measured in the presence of ammonia (NH3), humidity, and a number of volatile organic compounds at the chosen operating temperature of 55 degrees C" |
Keywords: | Adsorption Gold/*chemistry Mercury/*chemistry *Micro-Electrical-Mechanical Systems Particle Size *Quartz Crystal Microbalance Techniques *Sound Surface Properties; |
Notes: | "MedlineKabir, K M Mohibul Sabri, Ylias M Esmaielzadeh Kandjani, Ahmad Matthews, Glenn I Field, Matthew Jones, Lathe A Nafady, Ayman Ippolito, Samuel J Bhargava, Suresh K eng Comparative Study Research Support, Non-U.S. Gov't 2015/07/15 Langmuir. 2015 Aug 4; 31(30):8519-29. doi: 10.1021/acs.langmuir.5b01858. Epub 2015 Jul 23" |