Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractEvaluation of the Sealing Capability of the Implant Healing Screw by Using Real Time Volatile Organic Compounds Analysis: Internal Hexagon Versus Cone Morse    Next AbstractGeosmin suppresses defensive behaviour and elicits unusual neural responses in honey bees »

Front Biosci (Elite Ed)


Title:Evaluation of a new antibacterial coating of the internal chamber of an implant via real time measurement of Volatile Organic Compounds (VOCs)
Author(s):Scarano A; de Oliveira PS; Leo L; Festa F; Carinci F; Lorusso F;
Address:"Department of Innovative Technologies in Medicine & Dentistry, University of Chieti-Pescara, 66100 Chieti, Italy. Department of Oral Implantology, Dental Research Division, College Inga, UNINGA, 29312 Cachoeiro de Itapemirim, Espirito Santo, Brazil. Department of Morphology, Surgery and Experimental Medicine, University of Ferrara, 44121 Ferrara, Italy"
Journal Title:Front Biosci (Elite Ed)
Year:2021
Volume:13
Issue:2
Page Number:216 - 225
DOI: 10.52586/E879
ISSN/ISBN:1945-0508 (Electronic) 1945-0494 (Linking)
Abstract:"The aim of the present investigation was to evaluate the efficacy of an antibacterial coating of implant-abutment prosthetic junctions by real time measurement of Volatile Organic Compounds (VOCs). A total of 20 patients and 40 internal prosthetic junction implants were evaluated in the present investigation: 20 fixtures with antibacterial internal coating (Test) and 20 without treatment (Control). The VOCs measurements were evaluated at the baseline (T(0)) after the cover unit unscrewing, after 7 days (T(1)) and at 14 days (T(2)). No significant difference were detected at T0 (baseline), as Test and Control groups showed a VOCs max peak mean respectively of 2.15 +/- 0.71 and 2.21 +/- 0.69 (p > 0.05). At T(1) and T(2) as significant difference between the Test and Control Groups was detected (p < 0.01). At T2 the Test max peak was 2.29 +/- 0.73 and the Control was 3.65 +/- 0.91 (p < 0.01). The antibacterial internal coating demonstrated the capacity to prevent microbial VOCS activity at the level of the implant internal chamber and could be useful for long-term peri-implant tissue health"
Keywords:Anti-Bacterial Agents/pharmacology Humans *Volatile Organic Compounds Dental implant Peri-implant tissues Surface coating Titanium material Volatile Organic Compounds;
Notes:"MedlineScarano, Antonio de Oliveira, Pablo Santos Leo, Lucia Festa, Felice Carinci, Francesco Lorusso, Felice eng Singapore 2021/12/24 Front Biosci (Elite Ed). 2021 Dec 20; 13(2):216-225. doi: 10.52586/E879"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 04-12-2024