Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractEvaluation and predictive modeling of shelf life of minced beef stored in high-oxygen modified atmosphere packaging at different temperatures    Next AbstractBiocontrol of Rice Seedling Rot Disease Caused by Curvularia lunata and Helminthosporium oryzae by Epiphytic Yeasts from Plant Leaves »

Analyst


Title:Characterization of ion processes in a GC/DMS air quality monitor by integration of the instrument to a mass spectrometer
Author(s):Limero TF; Nazarov EG; Menlyadiev M; Eiceman GA;
Address:"Wyle Science, Technology and Engineering Group, 1290 Hercules Drive, Houston, TX 77058, USA. thomas.f.limero@nasa.gov"
Journal Title:Analyst
Year:2015
Volume:140
Issue:3
Page Number:922 - 930
DOI: 10.1039/c4an01800a
ISSN/ISBN:1364-5528 (Electronic) 0003-2654 (Linking)
Abstract:"The air quality monitor (AQM), which included a portable gas chromatograph (GC) and a detector was interfaced to a mass spectrometer (MS) by introducing flow from the GC detector to the atmospheric pressure ion source of the MS. This small GC system, with a gas recirculation loop for carrier and detector make-up gases, comprised an inlet to preconcentrate volatile organic compounds (VOCs) in air, a thermal desorber before the GC column, a differential mobility spectrometer (DMS), and another DMS as an atmospheric pressure ionization source for the MS. Return flow to the internally recirculated air system of the AQM's DMS was replenished using purified air. Although ions and unreacted neutral vapors flowed from the detector through Viton(R) tubing into the source of the MS, ions were not detected in the MS without the auxillary ion source, (63)Ni as in the mobility detector. The GC-DMS-MS instrument provided a 3-D measurement platform (GC, DMS, and MS analysis) to explore the gas composition inside the GC-DMS recirculation loop and provide DMS-MS measurement of the components of a complex VOC mixture with performance significantly enhanced by mass-analysis, either with mass spectral scans or with an extracted ion chromatogram. This combination of a mobility spectrometer and a mass spectrometer was possible as vapors and ions are carried together through the DMS analyzer, thereby preserving the chromatographic separation efficiency. The critical benefit of this instrument concept is that all flows in and through the thoroughly integrated GC-DMS analyzer are kept intact allowing a full measure of the ion and vapor composition in the complete system. Performance has been evaluated using a synthetic air sample and a sample of airborne vapors in a laboratory. Capabilities and performance values are described using results from AQM-MS analysis of purified air, ambient air from a research laboratory in a chemistry building, and a sample of synthetic air of known composition. Quantitative measures of a stand-alone AQM are disclosed for VOCs in the ppb to ppm levels with an average precision of 5.8% RSD and accuracy from 4% to 28% error against a standard method"
Keywords:"Air Pollutants/*analysis Chromatography, Gas/*methods Gases/*analysis Ions Tandem Mass Spectrometry/*methods Volatile Organic Compounds/*analysis;"
Notes:"MedlineLimero, T F Nazarov, E G Menlyadiev, M Eiceman, G A eng Research Support, U.S. Gov't, Non-P.H.S. England 2014/12/17 Analyst. 2015 Feb 7; 140(3):922-30. doi: 10.1039/c4an01800a"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 01-07-2024