« Previous AbstractEvaluation of a real-time method for monitoring volatile organic compounds in indoor air in a Japanese university    Next Abstract"Functional analysis of TraA, the sex pheromone receptor encoded by pPD1, in a promoter region essential for the mating response in Enterococcus faecalis" »

J Occup Health


Title:Comparison of sensor characteristics of three real-time monitors for organic vapors
Author(s):Hori H; Ishimatsu S; Fueta Y; Hinoue M; Ishidao T;
Address:"Department of Environmental Management, School of Health Sciences, University of Occupational and Environmental Health"
Journal Title:J Occup Health
Year:2015
Volume:20141121
Issue:1
Page Number:13 - 19
DOI: 10.1539/joh.14-0146-OA
ISSN/ISBN:1348-9585 (Electronic) 1341-9145 (Linking)
Abstract:"OBJECTIVE: Sensor characteristics and performance of three real-time monitors for volatile organic compounds (VOC monitor) equipped with a photo ionization detector (PID), a sensor using the interference enhanced reflection (IER) method and a semiconductor gas sensor were investigated for 52 organic solvent vapors designated as class 1 and class 2 of organic solvents by the Ordinance of Organic Solvent Poisoning Prevention in Japan. METHODS: Test vapors were prepared by injecting each liquid solvent into a 50 l Tedlar(R) bag and perfectly vaporizing it. The vapor concentration was from one-tenth to twice the administrative control level for all solvents. The vapor concentration was measured with the monitors and a gas chromatograph equipped with a flame ionization detector simultaneously, and the values were compared. RESULTS: The monitor with the PID sensor could measure many organic vapors, but it could not detect some vapors with high ionization potential. The IER sensor could also detect many vapors, but a linear response was not obtained for some vapors. A semiconductor sensor could detect methanol that could not be detected by PID and IER sensors. CONCLUSIONS: Working environment measurement of organic vapors by real-time monitors may be possible, but sensor characteristics and their limitations should be known"
Keywords:"Air Pollution, Indoor/analysis Chromatography, Gas Environmental Monitoring/*methods Flame Ionization Gases/analysis Humans Japan Occupational Exposure/prevention & control Solvents/classification Volatile Organic Compounds/*analysis Volatilization;"
Notes:"MedlineHori, Hajime Ishimatsu, Sumiyo Fueta, Yukiko Hinoue, Mitsuo Ishidao, Toru eng Comparative Study Research Support, Non-U.S. Gov't Australia 2014/11/26 J Occup Health. 2015; 57(1):13-9. doi: 10.1539/joh.14-0146-OA. Epub 2014 Nov 21"

 
Back to top
 