Title:Needle-trap device for the sampling and determination of chlorinated volatile compounds
Author(s):Dobrzynska E; Buszewski B;
Address:"Department of Chemical, Aerosol and Biological Hazards, Central Institute for Labour Protection-National Research Institute, Warsaw, Poland; Department of Environmental Chemistry and Bioanalytics, Faculty of Chemistry, Nicolaus Copernicus University, Torun, Poland"
Journal Title:J Sep Sci
Year:2013
Volume:20130904
Issue:20
Page Number:3372 - 3378
DOI: 10.1002/jssc.201300571
ISSN/ISBN:1615-9314 (Electronic) 1615-9306 (Linking)
Abstract:"A needle-trap device, with immobilized sorbent inside the syringe, coupled with GC-MS was applied for air sampling and determination of chlorinated volatile organic compounds such as dichloromethane, trichloromethane, and tetrachloromethane. The application of a needle trap packed with combination of three sorbents including Tenax TA, Carbopack X, and Carboxen 1000 resulted in detection limits of few pg for chlorinated volatile compounds and recoveries of 99.2-102.8%. The extraction and desorption parameters were optimized within the study. As a result, the precision determined as RSD was equal to 5.05 and 3.03 and 6.52% for dichloromethane, trichloromethane, and tetrachloromethane, respectively. The storage time for chlorinated compounds up to 48 h and reusability of the needle-trap device were verified. The obtained results have proved the ability of needle traps to compete with other solventless sampling and sample preparation extraction techniques"
Keywords:Carbopack X Carboxen 1000 Chlorinated compounds Needle trap Tenax TA;
Notes:"PubMed-not-MEDLINEDobrzynska, Elzbieta Buszewski, Boguslaw eng Germany 2013/08/21 J Sep Sci. 2013 Oct; 36(20):3372-8. doi: 10.1002/jssc.201300571. Epub 2013 Sep 4"

 