Title: | Analysis by desorption of volatile impurities from an ionic liquid solution in an unmodified gas chromatograph inlet |
Author(s): | Von Wald G; Albers D; McCabe T; |
Address: | "The Dow Chemical Company, Analytical Sciences, 1897 Building, Midland, MI 48667, USA. gvwald@dow.com" |
DOI: | 10.1016/j.chroma.2009.05.064 |
ISSN/ISBN: | 1873-3778 (Electronic) 0021-9673 (Linking) |
Abstract: | "A procedure for using ionic liquids to determine volatile impurities in compounds or matrices that are soluble in an ionic liquid is described. Using a conventional autosampler a droplet of ionic liquid solution is suspended in the inlet of the gas chromatograph and analytes are desorbed onto a GC column using splitless injection conditions. Results are presented for 1,2-propanediol, nonane, N,N-dimethylacetamide, and mesitylene in two different compounds in the ionic liquids trihexyltetradecylphosphonium bis(trifluoromethylsulfonyl)imide and trihexyltetradecylphosphonium dicyanamide" |
Keywords: | "Acetamides/analysis Alkanes/analysis Benzene Derivatives/analysis Chromatography, Gas/instrumentation/*methods Equipment Contamination Ionic Liquids/*chemistry Propylene Glycol/analysis Reproducibility of Results Volatile Organic Compounds/*analysis;" |
Notes: | "MedlineVon Wald, Grant Albers, David McCabe, Terry eng Netherlands 2009/06/30 J Chromatogr A. 2009 Jul 31; 1216(31):5927-30. doi: 10.1016/j.chroma.2009.05.064. Epub 2009 Jun 3" |