Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractCinnamomeoventrolide - Double Bond Regioisomerism in Frog Semiochemicals    Next AbstractThe life-cycle of the bont tick Amblyomma hebraeum in vitro »

J Environ Monit


Title:Design and field application of an automated cartridge sampler for VOC concentration and flux measurements
Author(s):Kuhn U; Dindorf T; Ammann C; Rottenberger S; Guyon P; Holzinger R; Ausma S; Kenntner T; Helleis F; Kesselmeier J;
Address:"Max Planck Institute for Chemistry, Biogeochemistry Department, Mainz, Germany. kuhn@mpch-mainz.mpg.de"
Journal Title:J Environ Monit
Year:2005
Volume:20050427
Issue:6
Page Number:568 - 576
DOI: 10.1039/b500057b
ISSN/ISBN:1464-0325 (Print) 1464-0325 (Linking)
Abstract:"One of the major limitations in advancing the understanding of tropospheric ozone and aerosol generation and developing strategies for their control is the technical ability to accurately measure volatile organic compounds (VOCs). This paper describes the design of a constant flow VOC sampler. The versatile sampler can be used for fully automated concentration and flux measurements of VOCs. The sampler incorporates a microprocessor control unit and provides highly accurate mass flow control and significant ease of operation. Sampling sequences can be programmed directly or by remote control through a PC. All important operational parameters necessary for a complete sampling audit trail are logged. Compact weatherproof housings and low power consumption allow operation at remote sites and locations which are sensitive to disturbances or have restricted access. Inner wetted surfaces of the sampler are constructed from non-contaminating materials that do not sorb or emit VOC, and thus permit the collection of representative samples even in environments with very low VOC concentrations. The cartridge magazine provides a maximum of 20 sequential cartridge samples, which allows for long-term air quality assessments. In the dual channel mode, two samples can be collected simultaneously through two independent sample loops, providing ten sequential sample pairs. This design allows the parallel collection of (a) quality assurance backup samples, (b) samples on two different types of cartridges/sorbents to allow a variety of analyses, or (c) differential samples for flux measurements using enclosure, aerodynamic profile, or relaxed eddy accumulation (REA) methods. Field applications including airplane profile measurements above a tropical rainforest area, as well as gradient and REA measurements over a mid-latitude mixed forest stand are described, and demonstrate the validity and flexibility of the system. In particular, the application of the VOC sampler as an integrated part of a REA system is emphasized"
Keywords:Adsorption Air Pollutants/*analysis Automation Environmental Monitoring/instrumentation/*methods Equipment Design Organic Chemicals/*analysis Reproducibility of Results Sensitivity and Specificity Time Factors Volatilization;
Notes:"MedlineKuhn, Uwe Dindorf, Tamara Ammann, Christoph Rottenberger, Stefanie Guyon, Pascal Holzinger, Ruppert Ausma, Sandra Kenntner, Thomas Helleis, Frank Kesselmeier, Jurgen eng Research Support, Non-U.S. Gov't England 2005/06/03 J Environ Monit. 2005 Jun; 7(6):568-76. doi: 10.1039/b500057b. Epub 2005 Apr 27"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 26-12-2024