Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractGeographical origin identification of two salmonid species via flavor compound analysis using headspace-gas chromatography-ion mobility spectrometry combined with electronic nose and tongue    Next AbstractOdorous gas emissions from sewage sludge composting windrows affected by the turning operation and associated health risks »

Sci Total Environ


Title:Improving the analytical flexibility of thermal desorption in determining unknown VOC samples by using re-collection
Author(s):Duan Z; Kjeldsen P; Scheutz C;
Address:"Department of Environmental Engineering, Building 115, Technical University of Denmark, 2800 Kgs. Lyngby, Denmark. Electronic address: duanzhe@env.dtu.dk. Department of Environmental Engineering, Building 115, Technical University of Denmark, 2800 Kgs. Lyngby, Denmark"
Journal Title:Sci Total Environ
Year:2021
Volume:20201230
Issue:
Page Number:144692 -
DOI: 10.1016/j.scitotenv.2020.144692
ISSN/ISBN:1879-1026 (Electronic) 0048-9697 (Linking)
Abstract:"The thermal desorption (TD) technique has long suffered from the 'one-shot' problem, whereby the entire sample is consumed in a single analysis, and thus no sample remains for repeated analysis. Recent developments in TD equipment allow for the quantitative re-collection of split samples during thermal desorption, which can be utilised for archiving or immediate analysis. However, the performance of TD systems for re-collecting different volatile organic compounds (VOCs) has rarely been demonstrated. This study provides a systematic investigation into the re-collection efficiency for over 90 VOCs on a TD unit under different conditions. An analytical method was developed based on multi-sorbent tubes and TD-GC/MS, which could quantitatively measure 92 VOCs with good sensitivity (method detection limit between 0.01 and 2 ng) and precision (< 10%). Satisfactory re-collection performance (recoveries within 100% +/- 20%) was found for over 70 compounds under different split modes for multiple times, and the single (outlet) split mode was preferred in this regard, in order to avoid significant uncertainties in the results. Thermal labile, polar or reactive compounds such as alcohols and ketones were generally not compatible with re-collection, as they were either lost due to thermal decomposition or formed as system artefacts. In addition, bromochloromethane should not be used as an internal standard when performing sample re-collection, since it will experience significant loss during repeated analysis and lead to overestimation for corresponding compounds. Finally, the re-collection was tested with low-concentration field samples to resolve the unexpected water problem in analysis. Although higher uncertainties were expected in the re-collected samples, the results provided good information on overall concentration variations at the sampling site, thereby instilling confidence in the results obtained from the primary analysis"
Keywords:Humidity Landfill trace gas analysis One-shot limitation Repeated analysis Sorbent tubes TD-GC-MS analysis Water interference;
Notes:"PubMed-not-MEDLINEDuan, Zhenhan Kjeldsen, Peter Scheutz, Charlotte eng Netherlands 2021/01/13 Sci Total Environ. 2021 May 10; 768:144692. doi: 10.1016/j.scitotenv.2020.144692. Epub 2020 Dec 30"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 25-12-2024