« Previous AbstractSynchronized discharge ionization for analysis of volatile organic compounds using a hand-held ion trap mass spectrometer    Next Abstract[VOCs Emission from Motor Vehicles in China and Its Impact on the Atmospheric Environment] »

J Chromatogr A


Title:Sub-second thermal desorption of a micro-sorbent trap for the analysis of ambient volatile organic compounds
Author(s):Chen TY; Li MJ; Wang JL;
Address:"Institute of Earth Sciences, Academia Sinica, Taipei 11529, Taiwan"
Journal Title:J Chromatogr A
Year:2002
Volume:976
Issue:1-Feb
Page Number:39 - 45
DOI: 10.1016/s0021-9673(02)01073-7
ISSN/ISBN:0021-9673 (Print) 0021-9673 (Linking)
Abstract:"This study investigates a novel approach of fast thermal desorption on a micro-sorbent trap for analyzing ambient volatile organic compounds (VOCs) by gas chromatography with flame ionization detection. Unlike conventional approaches, the temperature feedback mechanism for temperature control was abandoned, which often poses a limit to the heating speed due to slow response of the sensor and the control algorithm. Instead, a series of programmed a.c. pulses was given to the Ni-Cr wire coiled around the micro-trap to perform instant heating from room temperature to 250 degrees C within a fraction of a second, maintained at 250 degrees C during injection, and subsequently to 300 degrees C for trap cleaning. Temperature fluctuation around a high temperature set point could be maintained within +/- 10 degrees C. Significant improvement in resolution and peak height was obtained compared to a trap with temperature feedback and control algorithm. While keeping resolution at a satisfactory level, the sub-second desorption approach allows faster chromatography and at the same time increases the sensitivity of VOC analysis"
Keywords:"Chromatography, Gas/*methods Hot Temperature Organic Chemicals/*analysis Reproducibility of Results Volatilization;"
Notes:"MedlineChen, Tai-Yi Li, Ming-Jay Wang, Jia-Lin eng Research Support, Non-U.S. Gov't Netherlands 2002/12/05 J Chromatogr A. 2002 Nov 8; 976(1-2):39-45. doi: 10.1016/s0021-9673(02)01073-7"

 