Bedoukian   RussellIPM   RussellIPM   Piezoelectric Micro-Sprayer


Home
Animal Taxa
Plant Taxa
Semiochemicals
Floral Compounds
Semiochemical Detail
Semiochemicals & Taxa
Synthesis
Control
Invasive spp.
References

Abstract

Guide

Alphascents
Pherobio
InsectScience
E-Econex
Counterpart-Semiochemicals
Print
Email to a Friend
Kindly Donate for The Pherobase

« Previous AbstractDistance-dependent danger responses in bacteria    Next AbstractThe coordination of cell growth during fission yeast mating requires Ras1-GTP hydrolysis »

J Chromatogr A


Title:Evaluation of lifetime and analytical performance of gas chromatographic inlet septa for analysis of reactive semivolatile organic compounds
Author(s):Westland J; Organtini K; Dorman FL;
Address:"The Pennsylvania State University, 107 Whitmore Laboratories, University Park, PA 16802, USA"
Journal Title:J Chromatogr A
Year:2012
Volume:20120316
Issue:
Page Number:72 - 77
DOI: 10.1016/j.chroma.2012.03.030
ISSN/ISBN:1873-3778 (Electronic) 0021-9673 (Linking)
Abstract:"A comparison of septum degradation data obtained from a variety of standard GC septa, including the Thermogreen LB-2, AG3 (Green), Thermolite, BTO and the Merlin Microseal were evaluated. This study was conducted to determine analytical system lifetime between inlet maintenance based on chromatographic performance of reactive compounds as a function of septum aging through repeated injections. It is widely accepted that gas chromatographic septa degrade relative to the operating temperature of the injector and also based on the physical interaction with the syringe needle. This interaction may cause shedding of particles which then accumulate in the injection port liner, or possibly on the GC column in rare cases. This will likely lead to reactive compound breakdown in the injection port of flash-vaporizing inlets and thus reduce the amount of time that an instrument may be operated between inlet maintenance. Evaluations of the septa were based on the performance of the organochlorine compounds, endrin and 4,4'-DDT. These two compounds act as general indicators for the overall inertness of the GC injector, and their performance may be used to extrapolate to the performance of many other compounds. The average lifetime in terms of the number of injections for the studied septa was determined to range from 151 injections to nearly 2000 injections, depending on the septum. These values were found to be repeatable within the same septa material, or within class. The injection lifetime was found to be statistically different between different septa, however, indicating that not all septa are equivalent in terms of their durability to repeated injections. Additionally, some septa were found to predominately cause degradation of endrin, while others were found to cause reactivity of 4,4'-DDT, indicating that a difference in septa material may cause differences in what types of compounds become reactive in the GC inlet"
Keywords:"Chromatography, Gas/*instrumentation Volatile Organic Compounds/*analysis;"
Notes:"MedlineWestland, Jessica Organtini, Kari Dorman, Frank L eng Research Support, Non-U.S. Gov't Netherlands 2012/04/17 J Chromatogr A. 2012 May 25; 1239:72-7. doi: 10.1016/j.chroma.2012.03.030. Epub 2012 Mar 16"

 
Back to top
 
Citation: El-Sayed AM 2024. The Pherobase: Database of Pheromones and Semiochemicals. <http://www.pherobase.com>.
© 2003-2024 The Pherobase - Extensive Database of Pheromones and Semiochemicals. Ashraf M. El-Sayed.
Page created on 26-11-2024