Title:An Integrated Microfabricated Chip with Double Functions as an Ion Source and Air Pump Based on LIGA Technology
Author(s):Li H; Jiang L; Guo C; Zhu J; Jiang Y; Chen Z;
Address:"School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin 541004, China. lihua@guet.edu.cn. Guangxi Experiment Center of Information Science, Guilin 541004, China. lihua@guet.edu.cn. School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin 541004, China. longxingzu@126.com. School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin 541004, China. woshizhu.123@163.com. School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin 541004, China. 15107429521@163.com. School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin 541004, China. zhangwei@guet.edu.cn. School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin 541004, China. Livia_zw@163.com"
Journal Title:Sensors (Basel)
Year:2017
Volume:20170104
Issue:1
Page Number: -
DOI: 10.3390/s17010087
ISSN/ISBN:1424-8220 (Electronic) 1424-8220 (Linking)
Abstract:"The injection and ionization of volatile organic compounds (VOA) by an integrated chip is experimentally analyzed in this paper. The integrated chip consists of a needle-to-cylinder electrode mounting on the Polymethyl Methacrylate (PMMA) substrate. The needle-to-cylinder electrode is designed and fabricated by Lithographie, Galvanoformung and Abformung (LIGA) technology. In this paper, the needle is connected to a negative power supply of -5 kV and used as the cathode; the cylinder electrodes are composed of two arrays of cylinders and serve as the anode. The ionic wind is produced based on corona and glow discharges of needle-to-cylinder electrodes. The experimental setup is designed to observe the properties of the needle-to-cylinder discharge and prove its functions as an ion source and air pump. In summary, the main results are as follows: (1) the ionic wind velocity produced by the chip is about 0.79 m/s at an applied voltage of -3300 V; (2) acetic acid and ammonia water can be injected through the chip, which is proved by pH test paper; and (3) the current measured by a Faraday cup is about 10 pA for acetic acid and ammonia with an applied voltage of -3185 V. The integrated chip is promising for portable analytical instruments, such as ion mobility spectrometry (IMS), field asymmetric ion mobility spectrometry (FAIMS), and mass spectrometry (MS)"
Keywords:Liga air pump ion source ionic wind needle-to-cylinder;
Notes:"PubMed-not-MEDLINELi, Hua Jiang, Linxiu Guo, Chaoqun Zhu, Jianmin Jiang, Yongrong Chen, Zhencheng eng Switzerland 2017/01/06 Sensors (Basel). 2017 Jan 4; 17(1):87. doi: 10.3390/s17010087"

 